Medical information workstations (e.g., computers at doctor's and nurse's stations) must comply with GB 9706.1, YY 0505 (domestic), and IEC 60601-1-2 (international) EMC standards. As general IT equipment used in medical environments, the standards require them to meet the EMC standards of Information Technology Equipment (ITE) (e.g., GB 9254, CISPR 32) and also possess a certain level of immunity to interference. This ensures stable computer operation in electromagnetic environments with multiple medical devices, such as wards and nurse's stations, preventing crashes, data errors, or peripheral device (e.g., printers, barcode scanners) malfunctions, thus guaranteeing the continuous availability of the medical information system.