Medical intelligent warehousing systems (e.g., automated pharmacies, operating room consumable management cabinets) must comply with GB 9706.1, YY 0505 (domestic), and IEC 60601-1-2 (international) EMC standards. These systems include multi-axis servo/stepper motor driven robotic arms, conveyor belts, barcode scanners, industrial control computers, and network equipment. The standards require that the combined electromagnetic emissions generated by the coordinated operation of numerous motors do not affect the system's control network, scanning and identification, and data processing units. Furthermore, the entire system must operate stably within the hospital's information network environment and not become a source of interference.