Arthroscopic systems are used in minimally invasive joint surgery, and their EMC standards must comply with GB 9706.1, YY 0505 (domestic), and IEC 60601-1-2 (international). Their composition is similar to that of a laparoscopy system, but they may be used more frequently in conjunction with power and energy devices such as planers and radiofrequency ablation pens. The standards require that their camera system have stronger resistance to interference generated by the planer motor and radiofrequency energy, providing stable, flicker-free images, while the system itself does not interfere with the normal operation of these power devices.