Leakage current requirements:
BF type equipment: leakage current ≤10μA in normal state, and ≤50μA in single fault state; CF type equipment needs to pass a special test for cardiac current, with stricter leakage current limits.
Housing leakage current: all types of equipment must have a leakage current ≤500μA.
Insulation strength: insulation resistance: patient connection part ≥10MΩ (500VDC), non-patient part ≥5MΩ.
Withstand voltage test: apply 4kVrms/60s between input and output, leakage current ≤10mA, no breakdown or flashover.